{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2025-0683","assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","state":"PUBLISHED","assignerShortName":"icscert","dateReserved":"2025-01-23T18:11:20.770Z","datePublished":"2025-01-30T18:17:35.867Z","dateUpdated":"2025-02-12T20:41:36.789Z"},"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","product":"CMS8000 Patient Monitor","vendor":"Contec Health","versions":[{"status":"affected","version":"All versions"}]}],"credits":[{"lang":"en","type":"finder","value":"An anonymous researcher reported these vulnerabilities to CISA."}],"descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."}],"value":"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."}],"metrics":[{"cvssV4_0":{"Automatable":"NOT_DEFINED","Recovery":"NOT_DEFINED","Safety":"NOT_DEFINED","attackComplexity":"LOW","attackRequirements":"PRESENT","attackVector":"NETWORK","baseScore":8.2,"baseSeverity":"HIGH","privilegesRequired":"NONE","providerUrgency":"NOT_DEFINED","subAvailabilityImpact":"NONE","subConfidentialityImpact":"NONE","subIntegrityImpact":"NONE","userInteraction":"NONE","valueDensity":"NOT_DEFINED","vectorString":"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N","version":"4.0","vulnAvailabilityImpact":"NONE","vulnConfidentialityImpact":"HIGH","vulnIntegrityImpact":"NONE","vulnerabilityResponseEffort":"NOT_DEFINED"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]},{"cvssV3_1":{"attackComplexity":"HIGH","attackVector":"NETWORK","availabilityImpact":"NONE","baseScore":5.9,"baseSeverity":"MEDIUM","confidentialityImpact":"HIGH","integrityImpact":"NONE","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-359","description":"CWE-359 Exposure of Private Personal Information to an Unauthorized Actor","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert","dateUpdated":"2025-01-31T17:07:47.786Z"},"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"}],"solutions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.</p><p>Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's <a target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\">safety communication</a>.</p>\n\n<br>"}],"value":"Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's  safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."}],"source":{"advisory":"ICSMA-25-030-01","discovery":"EXTERNAL"},"title":"Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor","x_generator":{"engine":"Vulnogram 0.2.0"}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2025-01-31T15:35:10.389Z"},"references":[{"url":"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"},{"url":"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"},{"url":"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"}],"title":"CVE Program Container","x_generator":{"engine":"ADPogram 0.0.1"}},{"metrics":[{"other":{"type":"ssvc","content":{"id":"CVE-2025-0683","role":"CISA Coordinator","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"partial"}],"version":"2.0.3","timestamp":"2025-01-30T19:06:59.525559Z"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2025-02-12T20:41:36.789Z"}}]}}