{"dataType":"CVE_RECORD","dataVersion":"5.1","cveMetadata":{"cveId":"CVE-2023-1966","assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","state":"PUBLISHED","assignerShortName":"icscert","dateReserved":"2023-04-10T14:50:41.262Z","datePublished":"2023-04-28T18:06:51.431Z","dateUpdated":"2025-01-16T21:35:30.057Z"},"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","product":"iScan Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"4.0.0"},{"status":"affected","version":"4.0.5"}]},{"defaultStatus":"unaffected","product":"MiniSeq Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"2.0 "}]},{"defaultStatus":"unaffected","product":"iSeq 100","vendor":"Illumina ","versions":[{"status":"affected","version":"All versions"}]},{"defaultStatus":"unaffected","product":"MiSeq Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"4.0 (RUO Mode)"}]},{"defaultStatus":"unaffected","product":"MiSeqDx Operating Software","vendor":"Illumina ","versions":[{"status":"affected","version":"4.0.1 "}]},{"defaultStatus":"unaffected","product":"NextSeq 500/550 Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"4.0"}]},{"defaultStatus":"unaffected","product":"NextSeq 550Dx Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"4.0 (RUO Mode)"}]},{"defaultStatus":"unaffected","product":"NextSeq 550Dx Operating Software","vendor":"Illumina ","versions":[{"lessThanOrEqual":"1.3.1","status":"affected","version":"1.0.0 ","versionType":"custom"},{"status":"affected","version":"1.3.3 "}]},{"defaultStatus":"unaffected","product":"NextSeq 1000/2000 Control Software","vendor":"Illumina ","versions":[{"lessThanOrEqual":"1.4.1","status":"affected","version":"0","versionType":"custom"}]},{"defaultStatus":"unaffected","product":"NovaSeq 6000 Control Software","vendor":"Illumina ","versions":[{"lessThanOrEqual":"1.7 ","status":"affected","version":"0","versionType":"custom"}]},{"defaultStatus":"unaffected","product":"NovaSeq Control Software","vendor":"Illumina ","versions":[{"status":"affected","version":"1.8"}]}],"credits":[{"lang":"en","user":"00000000-0000-4000-9000-000000000000","value":"Illumina reported these vulnerabilities to CISA. "}],"descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"Instruments with Illumina Universal Copy Service v1.x and\nv2.x contain an unnecessary privileges vulnerability. An unauthenticated\nmalicious actor could upload and execute code remotely at the operating system\nlevel, which could allow an attacker to change settings, configurations,\nsoftware, or access sensitive data on the affected product.\n\n\n\n\n\n"}],"value":"Instruments with Illumina Universal Copy Service v1.x and\nv2.x contain an unnecessary privileges vulnerability. An unauthenticated\nmalicious actor could upload and execute code remotely at the operating system\nlevel, which could allow an attacker to change settings, configurations,\nsoftware, or access sensitive data on the affected product.\n\n\n\n\n\n"}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"PHYSICAL","availabilityImpact":"HIGH","baseScore":7.4,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"LOW","scope":"CHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-250","description":"CWE-250 Execution with Unnecessary Privileges","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert","dateUpdated":"2023-04-28T18:06:51.431Z"},"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01"},{"tags":["vendor-advisory"],"url":"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html"}],"source":{"discovery":"INTERNAL"},"title":"CVE-2023-1966","workarounds":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"\n\n<span style=\"background-color: rgb(255, 255, 255);\">Illumina recommends using the </span><a target=\"_blank\" rel=\"nofollow\" href=\"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html\">UCS Vulnerability Instructions Guide</a><span style=\"background-color: rgb(255, 255, 255);\">&nbsp;based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software. </span>\n\n<br>"}],"value":"\nIllumina recommends using the  UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software. \n\n\n"}],"x_generator":{"engine":"Vulnogram 0.1.0-dev"}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-02T06:05:27.096Z"},"title":"CVE Program Container","references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01","tags":["x_transferred"]},{"tags":["vendor-advisory","x_transferred"],"url":"https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html"}]},{"metrics":[{"other":{"type":"ssvc","content":{"timestamp":"2025-01-16T20:30:42.910219Z","id":"CVE-2023-1966","options":[{"Exploitation":"none"},{"Automatable":"no"},{"Technical Impact":"total"}],"role":"CISA Coordinator","version":"2.0.3"}}}],"title":"CISA ADP Vulnrichment","providerMetadata":{"orgId":"134c704f-9b21-4f2e-91b3-4a467353bcc0","shortName":"CISA-ADP","dateUpdated":"2025-01-16T21:35:30.057Z"}}]}}