{"containers":{"cna":{"affected":[{"product":"n/a","vendor":"n/a","versions":[{"status":"affected","version":"n/a"}]}],"descriptions":[{"lang":"en","value":"An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device."}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"NONE","baseScore":8.1,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AC:L/AV:A/A:N/C:H/I:H/PR:N/S:U/UI:N","version":"3.1"}}],"problemTypes":[{"descriptions":[{"description":"n/a","lang":"en","type":"text"}]}],"providerMetadata":{"dateUpdated":"2021-08-25T11:38:25.000Z","orgId":"8254265b-2729-46b6-b9e3-3dfca2d5bfca","shortName":"mitre"},"references":[{"tags":["x_refsource_MISC"],"url":"https://www.bbraunusa.com/en.htm"},{"tags":["x_refsource_MISC"],"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"cve@mitre.org","ID":"CVE-2021-33886","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"n/a","version":{"version_data":[{"version_value":"n/a"}]}}]},"vendor_name":"n/a"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device."}]},"impact":{"cvss":{"attackComplexity":"LOW","attackVector":"ADJACENT","availabilityImpact":"NONE","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"UNCHANGED","userInteraction":"NONE","vectorString":"CVSS:3.1/AC:L/AV:A/A:N/C:H/I:H/PR:N/S:U/UI:N","version":"3.1"}},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"n/a"}]}]},"references":{"reference_data":[{"name":"https://www.bbraunusa.com/en.htm","refsource":"MISC","url":"https://www.bbraunusa.com/en.htm"},{"name":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/","refsource":"MISC","url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-04T00:05:51.093Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://www.bbraunusa.com/en.htm"},{"tags":["x_refsource_MISC","x_transferred"],"url":"https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/"}]}]},"cveMetadata":{"assignerOrgId":"8254265b-2729-46b6-b9e3-3dfca2d5bfca","assignerShortName":"mitre","cveId":"CVE-2021-33886","datePublished":"2021-08-25T11:38:25.000Z","dateReserved":"2021-06-06T00:00:00.000Z","dateUpdated":"2024-08-04T00:05:51.093Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}