{"containers":{"cna":{"affected":[{"product":"SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A","vendor":"n/a","versions":[{"status":"affected","version":"Dana Diabecare RS, AnyDana-i, AnyDana-A  All versions prior to 3.0"}]}],"descriptions":[{"lang":"en","value":"In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-294","description":"AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2021-01-19T21:17:59.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","ID":"CVE-2020-27269","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A","version":{"version_data":[{"version_value":"Dana Diabecare RS, AnyDana-i, AnyDana-A  All versions prior to 3.0"}]}}]},"vendor_name":"n/a"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications lacks replay protection measures, which allows unauthenticated, physically proximate attackers to replay communication sequences via Bluetooth Low Energy."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294"}]}]},"references":{"reference_data":[{"name":"https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01","refsource":"MISC","url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-04T16:11:36.578Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2020-27269","datePublished":"2021-01-19T21:17:59.000Z","dateReserved":"2020-10-19T00:00:00.000Z","dateUpdated":"2024-08-04T16:11:36.578Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}