{"containers":{"cna":{"affected":[{"product":"BIGCOMPANYSOFT SOFTWARE PRODUCT","vendor":"n/a","versions":[{"status":"affected","version":"VC150 prior to Version 1.7.15"}]}],"descriptions":[{"lang":"en","value":"Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-74","description":"IMPROPER NEUTRALIZATION OF SPECIAL ELEMENTS IN OUTPUT USED BY A DOWNSTREAM COMPONENT CWE-74","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2021-01-08T15:17:09.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-007-01"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","ID":"CVE-2020-27260","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"BIGCOMPANYSOFT SOFTWARE PRODUCT","version":{"version_data":[{"version_value":"VC150 prior to Version 1.7.15"}]}}]},"vendor_name":"n/a"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"IMPROPER NEUTRALIZATION OF SPECIAL ELEMENTS IN OUTPUT USED BY A DOWNSTREAM COMPONENT CWE-74"}]}]},"references":{"reference_data":[{"name":"https://us-cert.cisa.gov/ics/advisories/icsma-21-007-01","refsource":"MISC","url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-007-01"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-04T16:11:36.335Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://us-cert.cisa.gov/ics/advisories/icsma-21-007-01"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2020-27260","datePublished":"2021-01-08T15:17:09.000Z","dateReserved":"2020-10-19T00:00:00.000Z","dateUpdated":"2024-08-04T16:11:36.335Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}