{"containers":{"cna":{"affected":[{"defaultStatus":"unaffected","product":"Smart Model 25000 Patient Reader","vendor":"Medtronic","versions":[{"status":"affected","version":"All versions"}]}],"credits":[{"lang":"en","type":"finder","value":"Sternum, based in Tel Aviv, Israel, discovered and initially reported these vulnerabilities to Medtronic."}],"descriptions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>Medtronic MyCareLink Smart 25000 is \n\n<span style=\"background-color: rgb(255, 255, 255);\">&nbsp;vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device</span>\n\n<br></p>"}],"value":"Medtronic MyCareLink Smart 25000 is \n\n vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device"}],"metrics":[{"cvssV3_1":{"attackComplexity":"LOW","attackVector":"ADJACENT_NETWORK","availabilityImpact":"HIGH","baseScore":8.8,"baseSeverity":"HIGH","confidentialityImpact":"HIGH","integrityImpact":"HIGH","privilegesRequired":"NONE","scope":"CHANGED","userInteraction":"REQUIRED","vectorString":"CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H","version":"3.1"},"format":"CVSS","scenarios":[{"lang":"en","value":"GENERAL"}]}],"problemTypes":[{"descriptions":[{"cweId":"CWE-122","description":"CWE-122 Heap-based Buffer Overflow","lang":"en","type":"CWE"}]}],"providerMetadata":{"orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert","dateUpdated":"2025-05-22T19:37:12.795Z"},"references":[{"tags":["x_refsource_MISC"],"url":"https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html"},{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01"}],"solutions":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>A firmware update to eliminates these vulnerabilities has been developed by Medtronic and is available by updating the MyCareLink Smartapp via the associated mobile application store. Upgrading to the latest v5.2 mobile application version will ensure the Patient Reader is also updated on next use. The user’s smart phone must be updated to the following operating system version for the patches to be applied: iOS 10 and above; Android 6.0 and above.</p><p>Medtronic has released additional <a target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\">patient focused information</a>: </p><p><a target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/xg-en/product-security/security-bulletins.html\">https://www.medtronic.com/xg-en/product-security/security-bulletins.html </a></p>\n\n<br>"}],"value":"A firmware update to eliminates these vulnerabilities has been developed by Medtronic and is available by updating the MyCareLink Smartapp via the associated mobile application store. Upgrading to the latest v5.2 mobile application version will ensure the Patient Reader is also updated on next use. The user’s smart phone must be updated to the following operating system version for the patches to be applied: iOS 10 and above; Android 6.0 and above.\n\nMedtronic has released additional  patient focused information https://www.medtronic.com/security : \n\n https://www.medtronic.com/xg-en/product-security/security-bulletins.html  https://www.medtronic.com/xg-en/product-security/security-bulletins.html"}],"source":{"advisory":"ICSMA-20-345-01","discovery":"EXTERNAL"},"title":"Medtronic MyCareLink Smart Heap-based Buffer Overflow","workarounds":[{"lang":"en","supportingMedia":[{"base64":false,"type":"text/html","value":"<p>In response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader:</p><ul><li>Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts.</li><li>Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior.</li></ul><p>Medtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should:</p><ul><li>Maintain good physical control over home monitors.<ul><li>This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.</li></ul></li><li>Use only home monitors obtained directly from your healthcare provider or a Medtronic representative.</li><li>Patients should ensure that the operating system of their mobile phone is updated to the latest version of the available Android or Apple iOS operating system.</li></ul><p>Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.</p>\n\n<br>"}],"value":"In response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader:\n\n  *  Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts.\n  *  Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior.\n\n\nMedtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should:\n\n  *  Maintain good physical control over home monitors.  *  This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.\n\n\n\n  *  Use only home monitors obtained directly from your healthcare provider or a Medtronic representative.\n  *  Patients should ensure that the operating system of their mobile phone is updated to the latest version of the available Android or Apple iOS operating system.\n\n\nReport any concerning behavior regarding these products to your healthcare provider or a Medtronic representative."}],"x_generator":{"engine":"Vulnogram 0.2.0"},"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","ID":"CVE-2020-25183","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"Medtronic MyCareLink Smart 25000 Reader","version":{"version_data":[{"version_value":"Smart 25000 all versions"}]}}]},"vendor_name":"n/a"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"Medtronic MyCareLink Smart 25000 all versions contain an authentication protocol vuln where the method used to auth between MCL Smart Patient Reader and MyCareLink Smart mobile app is vulnerable to bypass. This vuln allows attacker to use other mobile device or malicious app on smartphone to auth to the patient’s Smart Reader, fools the device into thinking its communicating with the actual smart phone application when executed in range of Bluetooth."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"IMPROPER AUTHENTICATION CWE-287"}]}]},"references":{"reference_data":[{"name":"https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01","refsource":"MISC","url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-04T15:26:10.249Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2020-25187","datePublished":"2020-12-14T19:18:52.000Z","dateReserved":"2020-09-04T00:00:00.000Z","dateUpdated":"2025-05-22T19:37:12.795Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}