{"containers":{"cna":{"affected":[{"product":"IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors","vendor":"ICS-CERT","versions":[{"status":"affected","version":"The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3"}]}],"datePublic":"2018-06-05T00:00:00.000Z","descriptions":[{"lang":"en","value":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory (\"write-what-where\") from an attacker-chosen device address within the same subnet."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-287","description":"IMPROPER AUTHENTICATION CWE-287","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2018-06-05T19:57:01.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","DATE_PUBLIC":"2018-06-05T00:00:00","ID":"CVE-2018-10597","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors","version":{"version_data":[{"version_value":"The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3"}]}}]},"vendor_name":"ICS-CERT"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory (\"write-what-where\") from an attacker-chosen device address within the same subnet."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"IMPROPER AUTHENTICATION CWE-287"}]}]},"references":{"reference_data":[{"name":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01","refsource":"MISC","url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-05T07:39:08.450Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2018-10597","datePublished":"2018-06-05T20:00:00.000Z","dateReserved":"2018-05-01T00:00:00.000Z","dateUpdated":"2024-09-16T23:46:22.088Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}