{"containers":{"cna":{"affected":[{"product":"Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump","vendor":"n/a","versions":[{"status":"affected","version":"Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump"}]}],"datePublic":"2018-02-15T00:00:00.000Z","descriptions":[{"lang":"en","value":"A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-120","description":"CWE-120","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2018-02-15T10:57:01.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"},{"name":"101252","tags":["vdb-entry","x_refsource_BID"],"url":"http://www.securityfocus.com/bid/101252"},{"name":"100665","tags":["vdb-entry","x_refsource_BID"],"url":"http://www.securityfocus.com/bid/100665"},{"name":"43776","tags":["exploit","x_refsource_EXPLOIT-DB"],"url":"https://www.exploit-db.com/exploits/43776/"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","ID":"CVE-2017-12718","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump","version":{"version_data":[{"version_value":"Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump"}]}}]},"vendor_name":"n/a"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"CWE-120"}]}]},"references":{"reference_data":[{"name":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A","refsource":"MISC","url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"},{"name":"101252","refsource":"BID","url":"http://www.securityfocus.com/bid/101252"},{"name":"100665","refsource":"BID","url":"http://www.securityfocus.com/bid/100665"},{"name":"43776","refsource":"EXPLOIT-DB","url":"https://www.exploit-db.com/exploits/43776/"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-05T18:43:56.539Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"},{"name":"101252","tags":["vdb-entry","x_refsource_BID","x_transferred"],"url":"http://www.securityfocus.com/bid/101252"},{"name":"100665","tags":["vdb-entry","x_refsource_BID","x_transferred"],"url":"http://www.securityfocus.com/bid/100665"},{"name":"43776","tags":["exploit","x_refsource_EXPLOIT-DB","x_transferred"],"url":"https://www.exploit-db.com/exploits/43776/"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2017-12718","datePublished":"2018-02-15T10:00:00.000Z","dateReserved":"2017-08-09T00:00:00.000Z","dateUpdated":"2024-08-05T18:43:56.539Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}