{"containers":{"cna":{"affected":[{"product":"Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.","vendor":"Abbott Laboratories","versions":[{"status":"affected","version":"All versions of pacemakers manufactured prior to August 28, 2017"}]}],"datePublic":"2017-08-29T00:00:00.000Z","descriptions":[{"lang":"en","value":"The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-287","description":"Improper authentication CWE-287","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2018-04-26T09:57:01.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"},{"name":"100523","tags":["vdb-entry","x_refsource_BID"],"url":"http://www.securityfocus.com/bid/100523"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","DATE_PUBLIC":"2017-08-29T00:00:00","ID":"CVE-2017-12712","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.","version":{"version_data":[{"version_value":"All versions of pacemakers manufactured prior to August 28, 2017"}]}}]},"vendor_name":"Abbott Laboratories"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"Improper authentication CWE-287"}]}]},"references":{"reference_data":[{"name":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01","refsource":"MISC","url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"},{"name":"100523","refsource":"BID","url":"http://www.securityfocus.com/bid/100523"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-05T18:43:56.613Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"},{"name":"100523","tags":["vdb-entry","x_refsource_BID","x_transferred"],"url":"http://www.securityfocus.com/bid/100523"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2017-12712","datePublished":"2018-04-25T13:00:00.000Z","dateReserved":"2017-08-09T00:00:00.000Z","dateUpdated":"2024-09-17T03:48:37.996Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}