{"containers":{"cna":{"affected":[{"product":"SIGMA Spectrum Infusion System","vendor":"Baxter","versions":[{"status":"affected","version":"6.05 (model 35700BAX) with wireless battery module (WBM) version 16"}]}],"datePublic":"2015-06-30T00:00:00.000Z","descriptions":[{"lang":"en","value":"Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes."}],"problemTypes":[{"descriptions":[{"cweId":"CWE-259","description":"Hard-coded password CWE-259","lang":"en","type":"CWE"}]}],"providerMetadata":{"dateUpdated":"2019-03-26T14:59:52.000Z","orgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","shortName":"icscert"},"references":[{"tags":["x_refsource_MISC"],"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01"}],"x_legacyV4Record":{"CVE_data_meta":{"ASSIGNER":"ics-cert@hq.dhs.gov","ID":"CVE-2014-5434","STATE":"PUBLIC"},"affects":{"vendor":{"vendor_data":[{"product":{"product_data":[{"product_name":"SIGMA Spectrum Infusion System","version":{"version_data":[{"version_value":"6.05 (model 35700BAX) with wireless battery module (WBM) version 16"}]}}]},"vendor_name":"Baxter"}]}},"data_format":"MITRE","data_type":"CVE","data_version":"4.0","description":{"description_data":[{"lang":"eng","value":"Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes."}]},"problemtype":{"problemtype_data":[{"description":[{"lang":"eng","value":"Hard-coded password CWE-259"}]}]},"references":{"reference_data":[{"name":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01","refsource":"MISC","url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01"}]}}},"adp":[{"providerMetadata":{"orgId":"af854a3a-2127-422b-91ae-364da2661108","shortName":"CVE","dateUpdated":"2024-08-06T11:41:49.088Z"},"title":"CVE Program Container","references":[{"tags":["x_refsource_MISC","x_transferred"],"url":"https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01"}]}]},"cveMetadata":{"assignerOrgId":"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6","assignerShortName":"icscert","cveId":"CVE-2014-5434","datePublished":"2019-03-26T14:59:52.000Z","dateReserved":"2014-08-22T00:00:00.000Z","dateUpdated":"2024-08-06T11:41:49.088Z","state":"PUBLISHED"},"dataType":"CVE_RECORD","dataVersion":"5.1"}